Tandem Connection Monitoring
Tandem Connection Terminating Equipment (TCTE) Testing
4 Configure the OmniBER to send a valid VC-n to the TCTE Source. For this
5 Check that there are no alarms indicated on the OmniBER front panel.
6 Check for correct TC-APId (access point identifier) as follows:
7 Press
8 Select the
9 Set the Receiver settings to match those of your system.
10 Set the TCM PATH field to
11 Select the
12 Select the MONITOR field and choose
128
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
example we have selected VC-4 mapping over STM-1.
If TCM is not enabled on the TCTE then the instrument STATUS line (at bottom
of the display) will indicate "TCM Loss of Multiframe (TC-LOM)", and the
Alarm Seconds results on the Results page will indicate TC-LOM alarms . The
absence of errors confirms that TCM is configured correctly.
and select
RECEIVE
MAIN SETTINGS
HIGH ORDER
ORDER selection is enabled when a TU mapping is selected.
OVERHEAD MONITOR
that the TC-APId matches the APId as expected from the TCTE, an example is
given on the following page.
.
SDH
folder.
or
LOW ORDER
folder.
, check on the display
TCM MESSAGE
. The LOW