Error Monitoring
Testing
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Tandem Connection Monitoring
Tandem Connection Terminating Equipment (TCTE) Testing
Method: Use the OmniBER Transmitter to generate errors in the system under test
and verify on the Receiver that the correct error result is obtained.
13 Press the
TRANSMIT
key and select
14 Select the
TEST FUNCTION
15 Set the TEST FUNCTION field to
16 Select an ERROR ADD TYPE and RATE, for this example we have chosen to
add B3 BIP errors at a rate of 1E-5, as shown below.
17 Press
to start the test. See "Test Timing " page 105 for advice on
RUN/STOP
setting test periods.
SDH
.
folder.
and select
SDH
ERRORS & ALARMS
.
129