Power and fuses

Power and fuses

Power supply

The instrument power supply is rated to operate with international standard single-phase power.
Specification: 90 - 264 VAC, Frequency 47 – 63 Hz, 240 Watts.

Fuses

The power inlet module houses 2 x 5A cartridge fuses, 5 mm diameter x 20 mm long, fast acting.
• 250 VAC (5 Amp)
When required, replacement of fuses is performed as follows:
1. Turn off mains power and unplug the cable from the IEC connector at the rear of the instrument.
2. Remove fuse cover.
3. Remove faulty fuse.
4. Check that the replacement fuse is the correct rating for your instrument (see "Instrument
properties"
5. Insert new fuse and replace fuse cover.
112
on page
11).