Kit Removal - Thermo Scientific Gibco CTS Rotea User Manual

Counterflow centrifugation system
Table of Contents

Advertisement

(continued)
Alarm or
Critical
message
alarm
Over-pressure
condition
Communication
error

Kit removal

CAUTION! Seal all fluid lines using a tube sealer or manual clamps prior to unlocking the door to
prevent fluid movement when the instrument pinch valves are released.
1. Press
pushbutton when it is illuminated to unlock the door.
CTS
Rotea
Counterflow Centrifugation System User Guide
Warning
Yes
Trigger instrument stop state
Yes
Instrument continues running
protocol, GUI indicates error
until communications are
restored
Chapter 3 Basic instrument operation
Instrument response
Kit removal
GUI message
Blockage detected
Protocol run has stopped.
Ensure manual clamps are
open and bags are connected
correctly. Press the stop button
to relieve pressure to Bag X.
Restart the protocol or select a
protocol step to start from.
Communication error
Ensure that the cable between
the laptop interface and
instrument is connected.
3
69