Chapter 3 Basic instrument operation
3
Kit preparation, loading and removal
6. Manually confirm that the Single-Use Kit configuration matches the image and that manual clamps
have been opened.
7. Click Confirm to continue.
CAUTION! Manual clamps are provided for use on input and output lines on the Single-
Use Kit to prevent fluid movement until the instrument Pinch Valves are engaged. However,
manual clamps also represent a significant process hazard if left in the closed state. Whilst
the instrument should detect this failure mode as an over-pressure fault, it will stop the
process potentially compromising the batch. Release all manual clamps prior to starting the
automatic sequence.
If required by the selected protocol:
8. Enter data inputs.
58
CTS
Rotea
Counterflow Centrifugation System User Guide