Chapter 4 OPC-UA interface
4
OPC-UA interface
Table 3 Variable set when changed (continued)
Variable
ProtocolEnded
DoorState
100
Type
Enum
The state indicates that the protocol has
stopped and the reason that it stopped. Press
the stop button or write the ProtocolState to
StopProtocol will clear any Errors.
• NotApplicable – Not Applicable
• CompletedSuccessfully – Protocol
completed successfully
• UserStoppedRun – The user or OPC-UA
stopped the protocol
• DoorUnlatchedError – The instrument has
raised a door error; the door failed to
close or has opened unexpectedly
• PumpMotorError – The instrument has
raised a pump motor fault
• CentrifugeMotorError -The instrument
has raised a centrifuge motor fault
• ValveError – The instrument has raised
a valve error. A valve may have failed to
close or open properly
• OutOfBalanceError – The instrument has
raised an out of balance error, the
centrifuge may be out of balance
• MoistureInBowlError – The instrument
has raised a moisture in bowl error, the kit
may be leaking
• OverPressureConditionError – The
instrument has raised an over pressure
error; excessive pressure has been
detected in one of the kit tubes
• UnknownError – The instrument is in an
unknown error state
Enum
• DoorOpened – The door is closed
• DoorClosed – The door is open
CTS
Rotea
Counterflow Centrifugation System User Guide