Chapter 4 OPC-UA interface
4

OPC-UA interface

OPC-UA interface
Commands OPC-UA to Rotea
Table 2 OPC-UA command list
RescanKit
SetGuiUserLevel
SetLocalProtocolFilePath
SetPlayPauseButtonEnable
SetProtocolFileContents
SetProtocolFileByLookupTableId
SetParameter1 through SetParameter20
96
application
Variable
Type
Boolean
Force a kit rescan:
• True – The instrument will rescan the kit
barcode.
• False – No action.
• GUIManagedLogin – Application login
Enum
is managed by the GUI, OPC-UA
commands are ignored in this mode.
• ViewUser – The MES has control of the
Rotea
the instrument status.
• AdminUser – The GUI is logged in as
an administrator and has control the
Rotea
are ignored in this mode.
Note: When OPC-UA enabled, a
SetGuiUserLevel command from the OPC-UA
server will over-ride the current instrument log-
in state.
String
Set the file name path of the protocol on the
laptop to be loaded to the Rotea
Enum
Enable the Play/Pause button on the Rotea
instrument.
• Disabled
• AsPerProtocol
• Enabled
String
Send the protocol file contents to the laptop to
be loaded to the Rotea
UInt32
Id of the protocol file in the OPC-UA look
up table. The lookup table is defined in
OPCUAProtocolList.yaml
Int32
The Metadata parameter data for the protocol.
The associated name of the parameter can be
found in variables ParameterName1 through
ParameterName20 after the protocol has been
loaded.
CTS
Rotea
Counterflow Centrifugation System User Guide
Values/Description
instrument and GUI just displays
instrument, OPC-UA commands
instrument.
instrument.