Table 3 Variable set when changed (continued)
Variable
ErrorRaised
KitId
CTS
Rotea
Counterflow Centrifugation System User Guide
Type
Enum
• NoError – There are no errors
• InternalError – An internal application
error has occurred
• FileNotFound – The
SetLocalProtocolFilePath file cannot be
found
• JsonParsingError – The protocol file is
corrupted
• FileCannotBeOpened – The protocol file
cannot be opened
• CmdFailed – Generic fail, the last
command issued did not succeed
• ParameterNotSet – A parameter has not
been set, protocol not loaded
• ParameterOutOfRange – A parameter is
out of range, protocol not loaded
• KitNotLoaded – The kit is not loaded,
protocol not loaded
• KitNotValid – The kit is not valid, protocol
not loaded
• InstrumentNotReady – Changing to
the specified. ProtocolState cannot be
performed from the current state
• RecoveryStepOutOfRange – The
ProtocolStepNumber is not a valid step
in the protocol
• ProtocolListDoesNotExist -
OPCUAProtocolList.yaml cannot be
found
• ProtocolListCannotBeOpen – Cannot
open
OPCUAProtocolList.yaml
• ProtocolListParsingError –
OPCUAProtocolList.yaml format
invalid
• ProtocolListIdDoesNotExist – The
LookupTableId cannot be found in
OPCUAProtocolList.yaml
String
The Kit id of the loaded kit
Chapter 4 OPC-UA interface
OPC-UA interface
Value/Description
4
101