Correct: Pressure Sensor tubing in
1
Incorrect: Pressure Sensor tubing out
2
CAUTION! The instrument cannot detect if the pump tubing or Tube Retainer have been correctly
positioned. Check prior to closing the door.
CAUTION! The tube will sit slightly proud of the channel in the region of the Pressure Sensor so that
the tube is compressed into the required shape when the door is closed. If not correctly located in the
channel, the tubing may be crushed when the door is closed.
CTS
Rotea
Counterflow Centrifugation System User Guide
Chapter 3 Basic instrument operation
Kit preparation, loading and removal
3
53