Chapter 3 Basic instrument operation
3
Run a protocol
(continued)
Alarm or
Critical
message
alarm
Centrifuge error
Valve error
Out of balance
condition
Moisture in bowl
68
Warning
Yes
Trigger instrument stop state
Yes
Trigger instrument stop state
Yes
Trigger instrument stop state
Yes
Trigger instrument stop state
Instrument response
CTS
Rotea
Counterflow Centrifugation System User Guide
GUI message
Centrifuge error
Protocol run has stopped.
Press the stop button on
the insturment, close manual
clamps and open and close
the instrument door to reset.
Restart the protocol or select
a protocol step to start from. If
error persists, abort the run.
Valve error
Protocol run has stopped.
Press the stop button on
the instrument, close manual
clamps and open and close
the instrument door to reset.
Restart the protocol or select
a protocol step to start from. If
error persists, abort the run.
Instrument out of balance
Protocol run has stopped.
Press the stop button to clear
the error, then restart the
protocol or select a step to
start from. If error persists,
abort the run.
Centrifuge leak detected
Protocol run has stopped.
Press the stop button to
clear the error or temporarily
override the moisture sensor if
safe to do so. If error persists,
abort the run.