Connection In The Control Panel For Cable L2; Connecting The Power Cable - Dentsply Sirona Heliodent Plus Service Manual

Hide thumbs Also See for Heliodent Plus:
Table of Contents

Advertisement

Dentsply Sirona
Service Manual Heliodent Plus
DX4
X200
62 15 102 D3507
D3507.076.01.15.02 03.2018
6.3.3

Connection in the control panel for cable L2

This change is only required:
if cable L2 is available.
L2
With DX1 boards up to serial no. 26 999
1. Insert the cable L2 into X103 on the DX4
2. Insert the cable L2, as shown, into the strain relief and firmly tighten
the strain relief on the DX4.
The cable L2 should now be in the strain relief in a position that is
turned as compared with the previous installation, so that the ferrite
core can be placed well in the housing.
6.3.4

Connecting the power cable

CAUTION
Observe the permissible nominal voltage range!
Connect 120 VAC with 1-phase connection
200 - 240 VAC with 1-phase or 2-phase connection, 50/60 Hz
1-phase connection
● X200.1: L1 (phase 1)
● X200.2: N' (Do not use!)
● X200.3: N
● X200.4: PE
6 Repair
6.3 Replacing PC board DX1
89