7 Reprocessing

36
Instructions for Use Midwest Tradition / Tradition Pro / Quiet-Air
7
Reprocessing
7.1

Automated cleaning

The cleaning equipment used must be approved by its manufacturer
for the cleaning of dental instruments.
For further details, refer to the operating instructions supplied with the
unit.
ü
The high-speed handpiece is conditioned with a cleaning device.
1.
Check whether the high-speed handpiece is clean after
conditioning under good lighting (min. 500 lux) and color
rendering index (min. 80 Ra).
2.
If they are still dirty, repeat the process.
Ä
For further processing, the high-speed handpiece should be
dry and free of residue.
3.
Blow the high-speed handpiece out with max. 3 bar.
67 27 155  D3673.201.10.02.02