Thermo Scientific Multidrop Combi nL User Manual
Q: Do you need specific plate adapters for the Multidrop Combi nL?
A: No, the Multidrop Combi nL has a fixed plate adapter and does not
use any external plate adapters.
Q: What is the material of the Multidrop Combi nL priming vessel?
A: The Multidrop Combi nL priming vessel (Cat. no. N07271) is made
of blue polypropylene (PP).
Q: Can the Multidrop Combi nL priming vessel be autoclaved?
A: Yes, it can be autoclaved.
Q: Can the Multidrop Combi nL reagent reservoir be autoclaved?
A: Yes, it can be autoclaved.
Q: Where can you see the Multidrop Combi nL internal software version
number?
A: When the instrument is powered on, the version number appears on
the top of the screen during the startup animation.
Device information under the Options menu also provides the internal
serial number.
Q: What is the orifice of the dispensing valve?
A: The orifice is 100 µm.
Q: Do you need to recalibrate the instrument when a dispensing valve is
changed?
A: No, you do not have to recalibrate the instrument, because the valve
performance is so consistent.
Q: I have an old spare valve which does not have any valve correction factor
enclosed. Can I still use it?
A: Yes, you can use it. Old spare valves do not have this correction
factor and the default value 0 is alright. If an old valve replaces the
newer version, the correction factor needs to be set to value 0.