X/Y Offset; Predispensing Volume - Thermo Scientific Multidrop Combi nL User Manual

Table of Contents

Advertisement

Predispensing
Thermo Fisher Scientific

X/Y offset

The default dispensing X/Y offset is 0.0/0.0 mm. You can change the x-
and y-axis offset values if the plate used is not the standard one specified
in the plate type list (see Table 4–3).
Use the Left and Right arrow keys to set the dispensing X/Y offset in
the x-axis and the Up and Down arrow keys to set it in the y-axis
direction. While you change the offset values, the dispensing valve head
moves * accordingly. The offset is set in 0.05 mm increments in both
the + and – directions. The maximum offset value depends on the
selected plate type.
Press
OK
adjustments are disregarded and the main view returns to the Settings
menu.
* y-axis = the dispensing valve head moves;
x-axis = the plate carrier moves
You can change the predispensing volume — the volume that is
automatically dispensed before the start of plate filling, if necessary. The
volume
default predispensing value is 100 nl (= 2 x 50 nl). Some more viscose
solutions used may sometimes require extra predispensing.
Go to the Settings menu. Select the Predispense (Prime) row.
Press
OK
. If the
button is not pressed within 60 seconds, the
OK
.
Thermo Scientific Multidrop Combi nL User Manual
Routine Operation
Protocol settings
47