Routine Operation
Instrument options
64
Thermo Scientific Multidrop Combi nL User Manual
2. The value of the factor can be changed with the Up and Down
arrows in the Liquid factor window. The default value is 1.000 and
when changed, the volume dispensed will be changed in proportion
to the factor. In this case the instrument dispenses 0.984 times
liquid compared to the value in the calibration curve.
3. Save the protocol.
Note The factor affects the whole calibration curve throughout the
volume range, and the volumes are changed, respectively, in the same
proportion as the check volume (Figure 4–22). The Liquid factor is,
however, protocol sensitive and must be saved with the protocol. The
liquid factor does not affect the calibration curve itself.
Figure 4–22. Calibration curve vs. factor and volume changes