Routine Operation
Dispensing parameters
Volume value box
Press the
button for 0.5 seconds. The Volume value box is now
OK
activated (bright red). Adjust the dispensing volume of the row where
the cursor is located using the Up and Down arrow keys.
Press
after selecting the desired volume.
OK
Use the Down arrow key and select Accept and press
after selecting
OK
the row volumes you wish.
You can press the
button to exit the volume selection without
STOP
making any changes.
Selecting columns and
You can first select columns of a microplate with a certain volume and
rows (Selected wells)
then select rows of the same microplate with another volume, or vice
versa. Where the selected columns and rows cross each other, the wells
will be filled with the row or column volume that was last selected when
such a protocol is executed.
Refer to the example plate in Figure 4–21 where the 100 nl row
selection overwrites the 50 nl column selection made first.
In general, be careful if you choose to carry out cross-dispensing where
selected wells are concerned.
41
Thermo Fisher Scientific
Thermo Scientific Multidrop Combi nL User Manual