Installation
Operational check
24
Thermo Scientific Multidrop Combi nL User Manual
Light indicator
Priming vessel –
with a crest
in the middle where the
priming liquid is aimed at
to avoid splashing and a
drain
Figure 3–14. Priming vessel inserted into place
2. Ensure that the priming vessel is inserted correctly into its slot
(Figure 3–14).
3. Pull the protective cover over the dispensing valve head.
4. Press
The pump starts and the instrument raises the pressure.
l
The light indicator blinks while the instrument regulates the
l
pressure.
The light is stable once the pressure has been reached.
l
Caution Unfastening the tubing while the pressure is on can cause a
hazardous situation, flood or spillage.
Check that there is always enough liquid in the reagent reservoir
l
and ensure that the filter is below the liquid level.
Check after each dispensing that the priming vessel is not filled up
l
if the tube drain is not used.
5. Press
The lifting mechanism is lowered and the dispensing valves are
l
opened as long as you press
.
PRESSURE
.
PRIME
Safety grooves –
designed to slope and prevent
fluid from entering the instrument
s
.
PRIME
Clamp
Thermo Fisher Scientific