Accuracy check

Thermo Fisher Scientific
You set the Liquid, for example, Water, in the Volume and Liquid
window of the Main menu. Refer to "Dispensing volume and liquid"
on page 36. Five liquids are factory set.
Dispense the selected volume, measure the weight of the dispensed
liquid and calculate the liquid factor. Refer to "Factor calculation" on
page 63.
Use the Up and Down arrow keys to select the factor for the liquid
used. The default settings for the liquid, the dispensing speed and the
dispensing volume are presented in the Factor window.
Use the Left arrow key to check the dispensing volume.
Use the Right arrow key to open the Quick Guidance window.
Press
.
OK
To perform an accuracy check and factor calculation (see "Factor
calculation" on page 63), follow the procedures below:
Go to the Options menu. Select the Liquid factor row.