Releasing the
transport lock
Thermo Fisher Scientific
Warning Ensure that the mains switch (Figure 2–4) at the left of the
back panel is in the OFF position. Also ensure that the mains supply
cable is disconnected.
The instrument has a transport lock (Figure 3–5).
Make sure the
red
transport lock is released before operating the
instrument.
Figure 3–5. Transport lock and transport lock tag present
1. Unscrew the two screws marked 1 and 2 of the transport lock with
the hexagonal screwdriver supplied (Figure 3–6) so that the track
mechanism is loosened.
Screw 1
Figure 3–6. Removing the transport lock
2. Remove the screws, the transport lock piece and the transport lock
tag from the screw marked 1 (Figure 3–6). Keep the tag for future
relocation or transportation of the instrument. The transport lock is
now released.
s
Thermo Scientific Multidrop Combi nL User Manual
Installation
Installation setups
Screw 2
19