Error and
warning codes
Error
Plate X position error.
Plate Y position error.
Z position error.
No more memory for storing
user data.
Not primed!
Missing prime vessel.
Protective cover not in place.
Pressure failure.
Thermo Fisher Scientific
Chapter 8
Troubleshooting Guide
Note Do not use the instrument if it appears that it does not function
properly.
s
When an error is detected, the current operation is terminated. After an
error, it is best to abort the current run and restart from the beginning
after the problem is fixed. The Multidrop Combi nL internal software
has the following error messages (Table 8–8).
Table 8–8. Error messages reported
Cause
Plate carrier cannot move
Plate carrier cannot move
Dispensing valve lifting mechanism
cannot move
Too many protocols saved
Predispensing undone
Priming vessel missing
Protective cover not in place
When tubes or plugs are not fastened
properly
Examples of a few error messages that appear in the Multidrop Combi
nL internal software are shown below.
Action
Power off/on (or contact service).
Power off/on (or contact service).
Power off/on (or contact service).
Delete protocols that are not used.
Press the
button.
PRIME
Insert the priming vessel.
Pull the protective cover over the
dispensing valve head.
Check that the tube attachments and caps
are properly tightened.
Thermo Scientific Multidrop Combi nL User Manual
91