Maintenance

How to refit the transport lock

How to refit the
transport lock
82
Thermo Scientific Multidrop Combi nL User Manual
8. Remove the instrument from the bag.
9. Clean the instrument using a mild detergent (e.g. dishwashing
liquid).
10. Remove any stains using 70% ethanol.
11. Disinfect the liquid path of the instrument with isopropanol, 10%
bleach (for example, sodium hypochlorite), or 70% ethanol. Last,
flush with distilled and filtered laboratory-grade water.
12. Wash and disinfect the reagent reservoir.
13. The priming vessel and the reagent reservoir can be autoclaved. All
the tubings can be autoclaved in the same bag with the reagent
reservoir.
14. After performing this decontamination procedure, enclose a signed
and dated Certificate of Decontamination both inside the transport
package and attached to the outside of the package (see Appendix A:
"Certificate of Decontamination").
1. Remove the
red
transport lock from the back panel of the
instrument (Figure 5–33).
Figure 5–33. Transport lock storage location
2. Adjust the dispensing valve lifting mechanism height and plate
carrier position to fasten screw 2 of the transport lock
(Figure 5–35). Then fasten screw 1 and the transport lock tag to the
dispensing valve head. The transport lock is now fastened
(Figure 5–35).