Routine Operation

Emergency situations

Emergency
situations
66
Thermo Scientific Multidrop Combi nL User Manual
2. If there is only a short break in dispensing, no washing or cleaning
of the tubes is necessary, depending of course on the liquid used.
For example, crystallizable buffer solution requires immediate
washing and cleaning of the tubes after use.
3. If there is a longer break, rinse the tubings with distilled and filtered
laboratory-grade water. Refer to the sections under "Regular and
preventive maintenance" on page 68.
Warning Before a prolonged standby, run the tubings dry to avoid
microbial growth from developing.
4. Switch the Multidrop Combi nL off by pressing the mains switch
(Figure 2–4) at the left of the back panel of the instrument into the
OFF position. The pressure is released.
Note To pause operation and hold the instrument on standby, just
push the protective cover in to release the pressure.
5. Wipe the instrument surfaces with a soft cloth or tissue paper
moistened with deionized distilled water, a mild detergent
(e.g. dishwashing liquid) or soap solution.
6. If you have spilled infectious agents on the dispenser, disinfect with
70% alcohol or some other disinfectant (see "Decontamination
procedure" on page 80).
In case there is any abnormal situation during operation, such as fluids
spilling inside the instrument, follow the steps below:
1. Switch OFF the instrument (Figure 2–4).
2. Unplug the instrument immediately from the power supply.
3. Carry out appropriate corrective measures. However, do not
disassemble the instrument.
4. If the corrective measures taken do not help, contact authorized
technical service or your local Thermo Fisher Scientific
representative.