Maintenance
Regular and preventive maintenance
70
Thermo Scientific Multidrop Combi nL User Manual
Figure 5–23. Removing the reagent filter
Place a disposable reagent basin (Cat. no. 9510027) or a 96-well
l
plate filled with detergent, for example, filtered 1% Micro-90
[International Products Corporation] or equivalent, under the
dispensing valves so that the liquid is in contact with the valves
(Figure 5–24).
Figure 5–24. Cleaning the tubings using a reagent basin
Press the
button for several minutes.
EMPTY
l
Place another disposable reagent basin or a 96-well plate filled with
l
distilled and filtered laboratory-grade water under the dispensing
valves so that the liquid is in contact with the valves
(cf. Figure 5–24).
Press the
button for several minutes.
EMPTY
l