Chapter 6 Technical Specifications; General Specifications; Table 6-7. General Specifications - Thermo Scientific Multidrop Combi nL User Manual

Table of Contents

Advertisement

specifications
Thermo Fisher Scientific
Chapter 6
Technical Specifications
General
Thermo Fisher Scientific reserves the right to change any specifications
without prior notice as part of our continuous product development
program. The general specifications are presented in Table 6–7.
Table 6–7. General specifications

General specifications

Overall dimensions
Weight (total)
Mains power supply
Power consumption
Heat dissipation
Fuses
Operating conditions (indoor use only)
Transportation conditions
Storage conditions
Autoclaving conditions of the priming
vessel and reagent reservoir
Control panel
Interface
Plate types
Dispensing volume range
Dead volume
Dispensing speed
Dispensing accuracy
Dispensing precision (CV%)
* Reagent recovery option available
355 mm (14 in.) (W) x 375 mm (14.8 in.) (D) x 220 mm (8.6 in.) (H)
9.6 kg (21.2 lbs.)
100–240 Vac, 50/60 Hz, nominal (operating range 90–256 Vac)
100 VA max.
341 BTU max.
None
+10°C to +40°C
-40°C to +70°C, packed in transport packaging
-25°C to +50°C, packed in transport packaging
Pressure: 1 bar
Temperature: 121°C
Time: 20 min
Keys: arrow keys,
,
OK
START
,
PRESSURE
TIP WASH
Serial RS-232C, USB 1.1 (2.0 compatible)
96, 384 and 1536-well plates
Standalone instrument
50–999 nl, in 1 nl increments
1.00–9.99 µl, in 10 nl increments
10.0–50.0 µl, in 100 nl increments
< 1.2 ml*
384-well plate: 50 nl in 6 seconds
1536-well plate: 50 nl in 21 seconds
< 1 µl: ± 5%
> 1 µl: ± 2%
CV ≤ 10%
50 nl:
0.5 µl:
CV ≤ 5%
1–10 µl: CV ≤ 4%
> 10 µl: CV ≤ 2%
Thermo Scientific Multidrop Combi nL User Manual
,
,
,
,
STOP
PRIME
EMPTY
SHAKE
FILLit Software
in 1 nl increments
1 µl in 8 seconds
1 µl in 27 seconds
,