Default Alarm Options - Teledyne GT Series User Manual

Hide thumbs Also See for GT Series:
Table of Contents

Advertisement

GT SERIES
1 INSTRUMENT, 7 APPLICATIONS

5.6. Default Alarm Options

Alarm Type
Latching?
LEL Warning
LEL (HI)
LEL (HIHI)
O
(LO)
2
O
(LOLO)
2
O
(HIHI)
2
CO Warning
CO (HI)
CO (HIHI)
CO (STEL)
CO (LTEL/TWA)
H
S Warning
2
H
S (HI)
2
H
S (HIHI)
2
H
S (STEL)
2
H
S (LTEL/TWA)
2
Low Battery
Zero Fault
Sensor Fault
Flow Fault
Calibration
Expired
Service Expired
36
Mutable?
Alarm disabled by default
N
Y
Y
N
Alarm disabled by default
Y
N
Y
N
N
N/A
Alarm disabled by default
Y
N
Y
N
Y
N
Alarm disabled by default
Alarm disabled by default
Y
N
Y
N
Y
N
Y
N/A
Y
N/A
Y
N/A
Y
N/A
Y
N/A
Y
N/A
Audible
Visual LED
Indication
Indication
High Pitch
4x LED Flashing
High Pitch
8x LED Ramping
High Pitch
8x LED Ramping
High Pitch
8x LED Ramping
N/A
N/A
High Pitch
8x LED Ramping
High Pitch
8x LED Ramping
High Pitch
8x LED Ramping
High Pitch
8x LED Ramping
High Pitch
8x LED Ramping
High Pitch
8x LED Ramping
Low Pitch
Fault LED Flashing
Low Pitch
Fault LED ON
Low Pitch
Fault LED Flashing
Low Pitch
Fault LED Flashing
Low Pitch
Fault LED Flashing
Low Pitch
Fault LED Flashing
Revision 10
67112