The monitor has five options to alert the user when calibration is due or is overdue:
1. Cal Due disabled - no date is displayed. The monitor does not require a valid calibration to
operate.
2. Cal Due enabled - if overdue, monitor is switched OFF.
3. Cal Due enabled - if overdue, message is displayed as shown in
Date, but after five seconds, the monitor warm-up continues automatically.
4. Cal Due enabled - if overdue, message is displayed as shown in
Overdue
and the user is required to acknowledge the message before continuing the warm-
up.
Note: This is the default setting.
If calibration is due, the user must either:
• Press and hold YES
• Press and hold NO
5. Cal Due enabled - if overdue, user can extend the due date by up to 31 days.
If calibration is within the "extended period", the user must still acknowledge that Calibration
has expired.
• Press and hold YES
• Press and hold NO
Off
is displayed.
67112
Revision 10
Figure 8: Calibration Due Date
button for monitor warm-up to continue.
button to switch OFF.
Figure 9: Calibration Overdue
button for monitor warm-up to continue.
button to reject extended period. The screen in