5.5.3 Sensor Fault (After Warm-up)

There are two types of sensor faults:
1. 'ZERO FAULT' and a spanner (wrench) symbol alternating with a zero reading, as illustrated in
Figure 50: Zero
Fault.
• Apply relevant test gas for two minutes and then remove it.
• Allow the display to return to zero.
• Switch monitor OFF and ON again.
• If the fault persists, return the monitor for service.
2. 'ZERO FAULT' and a spanner (wrench) symbol, alternating with a gas reading, as illustrated in
Figure 49: Zero Fault
• Leave monitor ON for 30 to 60 minutes.
• Switch it OFF and ON.
• If the fault persists, return the monitor for service.

5.5.4 Sample / Flow Fault

If a sample fault occurs, 'FLOW FAULT' alternates with the mode name, as shown in
Flow Fault
(CSM).
• Pump switches off (except Confined Space mode)
• Pump symbol disappears
• Fault LED tuns ON.
Check sample line, filter or probe for blockage. Clear blockage then restart the pump by pressing
the FLOW ACK
button.