Note: Bump Test mode will only display gas ranges configured in the monitor.
Multiple gas cylinders may be required to fully bump test the monitor.
Each gas cylinder used is referenced using a CalGas number.
For example:
• CalGas1 - Methane PPM
• CalGas2 - Methane LEL
• CalGas2 - Methane Vol
• CalGas4 - CO/H
1. The current CalGas selection number is highlighted. A different CalGas number can be
selected by pressing the UP
In the example in
Figure 75: Example CalGas
CalGas2.
2. To enter/edit the gas cylinder concentration, press and hold the NEXT
the gas value to be altered (as shown in
or DOWN
buttons as necessary to enter the correct gas value.
Note: Pressing and holding either of the buttons will rapidly increase/decrease the
gas value.
3. Once the monitor has been programmed with the actual gas cylinder content, press and hold
the OK
button to start the test. The screen shown in
be displayed.