4. If the content of the chemical filter is to be replaced, hold the filter assembly in the upright
position, as shown in
Unscrew the filter housing adaptor.
Figure 68: Filter Housing Adaptor Removal
5. Discard the filter housing content.
6. The chemical housing filter disc must be replaced if it is damaged or contaminated.
Using the flat end of a pencil, push the filter from the sample side, as shown in
Housing Filter
Removal, and discard it.
Insert new filter disc (part no. 67138) into filter housing until fully seated.
67112
Revision 10
Figure 67: Filter Removal
Figure 68: Filter Housing Adaptor
GT SERIES
1 INSTRUMENT, 7 APPLICATIONS
Removal.
Figure 69: