1. Using a pencil, push the cotton filter from the threaded end of the housing, as shown in
72: Cotton Filter
Removal, then remove it.
2. Replace the cotton filter if it is contaminated or damaged (part no. 10077).
3. Check the probe washers and replace if damaged (Part no. 12379).
4. Reattach all parts (reverse of removal procedure).
Note: Do not to overtighten the probe adapter or cotton filter housing.
67112
Revision 10
Figure 71: Cotton Filter Removal
Figure 72: Cotton Filter Removal