Alarm Level; Factory Default Settings; Custom Alarm Presets; Usb - Teledyne GAS LASER Operator's Manual

Table of Contents

Advertisement

GAS LASER

Alarm Level

Factory Default Settings

The instrument is preprogrammed
with two (2) distance-based alarm
level thresholds for the operator to
easily toggle between. The factory
default setting is 100' to alarm at
200 PPM-M. A quick press on the
button shown below will:
Set 50' distance alarm at
100 PPM-M
Set 100' distance alarm at
200 PPM-M

Custom Alarm Presets

The operator may customize the
alarm level threshold for the 100'
and/or 50' setting. The custom
settings
replace
the
defaults, and become the alarm
presets for future sessions.
To change the Alarm Detection
Threshold:
1. Press (long press) the MENU
button
2. Press the down arrow to scroll
through the list and highlight

Alarm Level.

3. Press SELECT.
4. Select 50' or 100'.
5. Press the down arrow to
increase the number.
6. Press SELECT to advance
to the next digit.
7. Press BACK when done
with settings.

USB

The Gas Laser instrument is equipped two USB settings:

Command Mode

Select to use the USB port as a
communications port.

Bluetooth (BLE)

The Gas Laser instrument is equipped with Bluetooth 4.2 BLE to support future features and mobile applications.
Doc No. 16162 Rev A, GAS LASER Operator Manual
NOTICE
Alarm level threshold setting is shown on the screen.
factory
(50' selection shown above, 100' selection shown below.)

File Mode

Select to use the USB port for file
access as a flash drive. This is the
instrument's default mode.
NOTICE: A higher Alarm Level setting will require a
higher concentration of methane to alarm.
A quick press on the
between the two (2) distance-based presets.
Step 2-4
18