ImageXpress Pico Automated Cell Imaging System Installation Guide
To load the environmental control cassette:
1. If the instrument is powered off, press the Power button on the front of the instrument to
turn on the power. See
2. Press the button at the top right of the instrument to open the top door.
3. If the sample stage already contains a labware holder, remove it.
4. Insert the environmental control cassette (with nothing inside) in the sample stage from the
left end so that the contacts on the cassette meet the contacts on the sample stage.
5. Confirm that the environmental control cassette is firmly seated.
6. Lift the top of the environmental control cassette and insert the plate containing your
samples.
7. Replace the top of the environmental control cassette.
8. Press the button at the top right of the instrument to close the top door.
62
Powering On the Instrument on page 33
Note: The top door may not close if the cassette is not seated properly.
Note: The 4-slide holder is not compatible with the environmental control cassette.
for details.