ImageXpress Pico Automated Cell Imaging System Installation Guide

Protective Housing

The instrument features a protective outer housing, which is designed to protect you from
exposure to LED light, hot surfaces, and moving parts.
WARNING!
attempt to gain access to the interior of the instrument. These actions can damage the
instrument components and result in hazardous exposure to LED light, hot surfaces,
moving parts, or high voltage.

Electrical Safety

To prevent electrically related injuries and property damage, inspect all electrical equipment
before use and immediately report all electrical deficiencies. Contact Molecular Devices
Technical Support to service equipment that requires the removal of covers or panels.
WARNING! HIGH VOLTAGE.
shock hazard existing from a high-voltage source. Read and understand all safety
instructions before you install, maintain, or service the instrument.
Do not remove the instrument covers. To prevent electrical shock, use the supplied power
cords only and connect to a properly grounded wall outlet.
To ensure sufficient ventilation and provide access for disconnecting power from the
instrument, maintain a clearance of 20 cm to 30 cm (7.9 in. to 11.8 in.) at the rear of the
instrument.
If the instrument does not power on, you may need to replace the instrument fuses. See
Replacing Fuses on page 9
8
Do not manually open the top door, open the protective housing, or
Within the instrument is the potential of an electrical
for details.
5063666 K