1 1 0
.
+
1 3 0
In the Applic./Applic.3/CLASS./Param.2/OUTP.ACT menu, menu item 4.7,
you can make the following settings for the control outputs:
– switched off
– always on
– on at stability
The "SET" output normally changes its voltage level when the current weight
exceeds the minimum load. Alternatively, you can assign the "Ready for use"
function to this port.
Setting: Applic./Applic.3/CLASS./PARAM.2/CTRL.SET menu item 4.3.
Example 2: There should be three classes.
Configuration: The "Classification" application is selected, and printout has been
set up.
O
t Enter the class limits using the O key.
.
t E nter the upper limit for Class 1 using the keypad (in this example, 110 g)
O
t Save the upper limit for Class 1.
t E nter the upper limit for Class 1 using the keypad (in this example, 110 g).
O
t Save the upper limit for Class 1.
t P lace the sample on the weighing platform.
Operating Instructions Combics Indicators
Operation
83