Sartorius Combics Series Operating Instructions Manual page 74

Hide thumbs Also See for Combics Series:
Table of Contents

Advertisement

Available parameter settings * = Factory setting
A tare value entered manually overwrites a stored tare value (weight value). If you
enter a tare value manually, a tare value (weight value) stored later overwrites the
manually entered value.
Setting: Applic. / TARE.FNC menu item 3.25.2
Restore factory default settings: Applic./Reset menu item 9.1.
Target Checkweighing entails comparing the current weight value to a defined target. You
can enter the value for this target using the keypad, or by saving the weight value
indicated. You can also define upper and lower tolerance limits based on this target.
You can do this by:
– E ntering absolute values using the keypad or placing the desired amount of
weight on the platform and saving the value
or
– b y entering each value as a percentage deviation of the target weight
or
– b y entering each value as an asymmetrical percentage deviation of the target
weight that is selected via the keypad or using the r key
or
– b y entering a relative weight deviation from the target weight via the keypad.
The value remains valid until deleted by pressing the c key or until overwritten by
a new value. It remains saved after the scale is switched off.
Preparation t Open the APPLIC. / APPLIC.2 / CHECK.WG menu.
MIN.INIT Minimum load for initialization
Aut.strt Automatic start of applications when you switch on
the device with most recently saved application data
TARE.FNC Tare function
CHECK.RG Checkweighing range
1 digit
1 digit
2 digits
2 digit
5 digits
5 digit
10 digits
10 digT.
20 digits
20 digT.
50 digits
50 digT.
100 digits
100 diG.
200 digits
200 diG.
500 digits
500 diG.
1000 digits
1000 dI.
Automatic (on)
automat
Manual (off)
manual
C an add a preset tare if tare value is avail-
normal
able; however no tare function possible
3.25.1*
W hen a preset tare is entered, the tare
special
value is deleted; however, tare function
activation is possible
30 to 170%
30- 170%
10% to infinity
10-max.l
Operating Instructions Combics Indicators
Operation
3.5
3.5.1*
3.5.2
3.5.3
3.5.4
3.5.5
3.5.6
3.5.7
3.5.8
3.5.9
3.5.10
3.8
3.8.1*
3.8.2
3.25
3.25.2
4.2.
4.2.1*
4.2.2
73

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Combics 1Combics 2Caisl1Caisl2Cais1Cais2

Table of Contents