..., 50%, 100%), configured in Setup under: Activity menu item 3.19.
If the "Averaging" parameter is set to 2%, for example, and the animal or object
weighs 10 kg, measurement does not begin until the fluctuation in weight value
remains below 200 g during three consecutive measurements.
Display A calculated average value is shown continuously on the main display. The G symbol
indicates the calculated value.
You can toggle between the results display and the current scale display by pressing
the w key.
Setting: Applic./Applic.1/ANIM.WG/DIS.UNLD menu item 3.21.
You can select "Display is static until unload threshold reached" to have the display
switch automatically to the weight readout when you unload the weighing platform
(i.e., when the load is less than half the minimum load). The result of the most recent
averaging operation is not saved.
If you select "Display is static until the c key is pressed," the calculated average
remains displayed even after the weighing platform is unloaded, until you press the
c key to begin a new measurement.
Example: The weight of one mouse should be measured.
Configuration: The "Animal Weighing" application is selected, and printout has been
set up (see "Configuration").
t Place empty container on the scale.
)
t T are the scale.
This is not required if the automatic tare function is active. The tare weight is
saved automatically when you place the container on the platform.
t Place the mouse in the container.
2 0
t E nter the number of sub-weighing operations using the keypad (in this example,
20 measurements).
O
t Save the value entered and start the averaging.
Operating Instructions Combics Indicators
Operation
65