Operation
Example:
External calibration and manual adjustment with default weights (weighing parameters: factory settings)
(
J
Example 2
External calibration and manual adjustment with freely configurable adjustment weight (within the limits of 1/3 max.
(
J
48
Operating Instructions Combics Indicators
1.) Zero the scale.
2.) S tart calibration
(e.g., when adjustment
prompt flashes WP).
c al.Ext. is displayed for
two seconds.
Y ou are prompted to place
the required weight on the
platform (e.g., 10,000 g).
3.) P lace the calibration/-
adjustment weight on the
weighing platform.
1.) Set scale to zero.
2.) Start calibration (e.g.,
when adjustment prompt
flashes WP).
"cal.ext." is displayed
for two seconds.
Ext.
calibration
Nom.
+
10000 g
Diff. +
1 g
J
--------------------
24.02.2012
10:15
Type
CAISL2
Ser.no.
12345678
Vers.
1.0103.11.2
BVers.
01-26-02
--------------------
Ext.
calibration
Tar
+
10000 g
Diff. +
1 g
Ext.
adjustment
Diff. +
0 g
--------------------
24.02.2012
10:15
Name:
--------------------
5
)
+
T he difference between
the measured value and the
true weight of the sample
will be displayed with plus/
minus signs.
A printout will be
generated if the process
is cancelled using the (
4.) A ctivate calibration adjust-
ment (press the ( key to
cancel).
T he adjustment weight
is displayed once adjust-
ment is finished.
A GMP-compliant
printout is generated.
capacity to max. capacity)
You are prompted to place the
calibration/adjustment weight on
the platform (e.g., 10,000 g).
3.) Enter the desired CAL weight
above the number block and
confirm with ")" (5,000 g here).
If the weight is too heavy or
too light, an error message will
appear.