Configuring Weighing Platforms
Installation Location in Germany An exception to this is the setting for "Germany (Zone D):" If during external
Setup information – I t is only possible to enter geographical data when the menu access switch is
24
Operating Instructions Combics Indicators
The A/D converter can now be treated like a standard weighing platform in
connection with the load sensor.
t C lose the menu access switch (right position) and reattach the cap.
Once ADC configuration has been completed, an adjustment of the weighing
h
platform (calibration/adjustment and linearization) must be carried out (see page 47
and "Calibration/Adjustment without Weights", page 30).
E ntering Geographical Data for Use in Legal Metrology
Purpose Entering geographical data allows the external adjustment of weighing equipment
at a place (e.g. at the manufacturer or vendor's place of business) that is not the
same as the place of installation. If the weighing equipment is adjusted at the place
of installation, it is not necessary to enter geographical data.
The sensitivity of weighing equipment changes depending on the place of
installation as it is dependent on the on-site gravitational force – or, more precisely,
on gravitational acceleration. Saving geographical data makes it possible to change
the place of installation of the weighing equipment after external adjustment has
been carried out.
The adjustment of weighing equipment is valid at the place of installation and
within a specific tolerance zone. At 3000 e this zone extends ±100 km from the set
geographical latitude and ±200 m from the set elevation above sea level.
adjustment of weighing equipment within Germany the geographical data
– Geographical latitude: 51.00 degrees N
– 513 m elevation above sea level
are entered, the weighing equipment can be used throughout Germany.
Gravitational acceleration for "Germany (Zone D)" is 9.810 m/s
geographical data for "Germany (Zone D)" are entered in the output device.
It is recommended to use the geographical data settings for "Germany (Zone D)"
when adjusting and delivering the weighing equipment within Germany. Entering
exact geographical data will lead to a higher level of accuracy but will also restrict
the tolerance zone.
open.
– W hen the Service mode is activated, you can enter geographical data in the
Setup menu under "WP-1" for the first weighing platform and COm1 / WP-2,
UNICOM / WP-2 or COm-WP for the second weighing platform. The settings are
made in the corresponding Setup menu under menu item 1.20.
– Y ou can enter either the "geographical latitude in degrees" (LATITUD menu
item 1.20.1) and "elevation in m above sea level" (ALTITUD menu item 1.20.2),
or the value for gravitational acceleration (GRAVITY menu item 1.20.3).
G ravitational acceleration takes precedence over the geographical latitude and
elevation of the location: If it has been entered, input fields for latitude and
elevation show the values 99999.99 and 9999999 respectively. If only elevation
and latitude have been entered, 0000000 is displayed for gravitational
acceleration.
I f you return to the highest level of the Setup menu without saving the configura-
3
tion parameter beforehand (save menu item 1.20.4) any settings that have been
made will be deleted.
. On delivery the
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Combics 1Combics 2Caisl1Caisl2Cais1Cais2

Table of Contents