Operation
mDef
+
20
T
+
0.292 kg
x-Net +
0.183 kg
--------------------
66
Operating Instructions Combics Indicators
or
r
t S et the number of reference parts using r: 1, 2, 5, 10, 20, etc.
(
t Start the calculation of the reference piece weight.
The averaging routine does not begin until the fluctuation in weight value remains
below a defined threshold over three consecutive measurements. The number of
sub-weighing operations remaining is shown in the numeric display.
t T he averaging result is displayed.
p
t Print the results.
Note: If automatic printout of results is enabled, you do not need to press the
p key. The results are printed automatically.
Printout configuration: see page 96.
y W hen you unload the weighing platform, the display switches to the weight
readout automatically, unless configured otherwise in the menu. The weighing
instrument is ready for the next measurement.