Sartorius Combics Series Operating Instructions Manual page 65

Hide thumbs Also See for Combics Series:
Table of Contents

Advertisement

Operation
Available parameter settings * = Factory setting
( (
Minimum load for initialization You can set the minimum load here, i.e. the load that must be placed on the
Starting the Measurements The averaging routine does not begin until the fluctuation in weight value remains
64
Operating Instructions Combics Indicators
MIN.INIT Minimum load for initialization
1 digit
2 digit
5 digit
10 digT.
20 digT.
50 digT.
100 diG.
200 diG.
500 diG.
1000 dI.
START Start of averaging
manual
automat
ACTIVTY Animal activity
0.1 perc.
0.2 perc.
0.5 perc.
1 perc.
2 perc.
5 perc.
10 perc.
20 perc.
50 perc.
100 perc.
PRINT Autom. printout of results
manual
automat
DIS.UNLD Static display of result after load removed
CLEARED
PRESENT
)
t T o save the setting, press the ) key.
... t T o exit setup: Press the ( key several times.
weighing platform in order to carry out the application. If the load on the platform is
too light, the following will occur:
– Error code Inf 29 appears,
– The weighing platform is not initialized,
– The preset reference sample quantity is saved.
Setting: Applic./Applic.1/ANIM.WG/Min.init menu item 3.6.
The minimum load can be set in 10 steps from 1 to 1000 digits (see available
parameters). The "digits" here refer to the scale intervals for the connected weighing
platform. If the interval of the connected platform is 1 g, for example, and 1000
digits are required, you must place at least 1000 g (= 1000 intervals =1000 digits)
on the weighing platform for initialization.
below a defined threshold over three consecutive measurements. The tolerance limit
is defined as a percentage of the animal or object weight (for example, 0.1%, 0.2%,
1 digit*
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Manual*
Automatic
0.1% of animal/object
0.2% of animal/object*
0.5% of animal/object
1% of animal/object
2% of animal/object
5% of animal/object
10% of animal/object
20% of animal/object
50% of animal/object
100% of animal/object
Off*
On
Display is fixed until unload
threshold reached*
Fixed display until c is pressed
3.6
3.6.1
3.6.2
3.6.3
3.6.4
3.6.5
3.6.6
3.6.7
3.6.8
3.6.9
3.6.10
3.18.
3.18.1
3.18.2
3.19.
3.19.1
3.19.2
3.19.3
3.19.4
3.19.5
3.19.6
3.19.7
3.19.8
3.19.9
3.19.10
3.20.
3.20.1
3.20.2
3.21.
3.21.1
3.21.2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Combics 1Combics 2Caisl1Caisl2Cais1Cais2

Table of Contents