Configuring the A/D converter (ADC)
The weighing platform must already be connected.
Open the menu access switch
The menu access switch is located on the back of the indicator right next to
the weighing platform connection.
t Remove the cap.
t S lide the switch to the left (= "open" position).
e
t Switch off and restart the device.
t While all segments are lit, briefly press the ( key.
y ADC-CON appears briefly on the display and the S-CODE
y The cursor flashes on the display.
t Enter the service password (see Appendix).
)
t Confirm your entry using the ) key.
y T he device is in Service mode. This can be recognized by the small S in
the top right of the display.
t S elect the weighing platform to be configured, using the k key to
switch to WP-2 if required.
)
t Confirm your selection using the ) key.
t S elect the Configuration mode using the k key: STANDRD or VERIF..
t C arry out ADC configuration (see menu tree starting on p. 19).
t O nce you have completed the configuration, save the data using the
SAVE menu item.
y The indicator will restart automatically.
Configuring Weighing Platforms
Operating Instructions Combics Indicators
23