Setup information – A djustment without weights is only possible when the menu access switch is
30
Operating Instructions Combics Indicators

Adjustment Without Weights

In the Service menu, adjustment without weights can be carried out by entering the
characteristic data of the load cells.
3
Adjustment without weights may not be carried out on weighing equipment used
in legal metrology.
open in the Service menu.
– W hen the Service mode is activated, you can enter the necessary parameters for
adjustment without weights in the Setup menu under "WP-1" for the first
weighing platform and COm1 / WP-2, UNICOM / WP-2 or COm-WP for the second
weighing platform( only Combics2 ). The settings are made in the corresponding
Setup menu under menu item 1.19.
– T he "Nominal load" parameter must be entered in the kg unit.
– T he "Resolution" parameter must be entered in the kg unit and must correspond
to the scale interval d entered for the ADC configuration
– T he "Sensitivity" parameter is entered in mV/V (see the data sheet for the value).
The data entered are saved by selecting menu item 1.19.7. After saving, the data will
h
no longer be able to be read.
Procedure
t O pen menu access switch.
t A ctivate the Service mode.
t S elect the weighing platform.
t E nter the nominal load of the load cell(s) in kg under menu item 1.19.1. If the
weighing platform has multiple load cells, the nominal load must be multiplied
accordingly (e.g. 4 load cells, each of which has a capacity of 50 kg, will produce
a nominal load of 200 kg)
t E nter the resolution in kg under menu item 1.19.2. The value must correspond
to the scale interval d entered under menu item 11.4.1
t E nter the sensitivity of the load cells in mV/V under menu item 1.19.3.
For weighing platforms with multiple load cells: Enter the individual values of
the load cells in 1.19.3 to 1.19.6 or enter the average of all load cells in 1.19.3.
t Save the values for adjustment without weighing under menu item 1.19.8.
t Close the menu access switch.
Menu structure for adjustment without weights
ADJ.W/O.W Adjustment without weights (entering the characteristic data of the load cell(s))
NOM.LOAD Nominal load
RESOLUT Resolution
SENSIT.1 Sensitivity in mV/V for cell 1 (or average value for all load cells)
SENSIT.2 Sensitivity in mV/V for cell 2
SENSIT.3 Sensitivity in mV/V for cell 3
SENSIT.4 Sensitivity in mV/V for cell 4
ZER.POIN Zero point
SAVE Save values for 1. 19
only for older ADCs).
(
only for older ADCs).
(
1.19.
1.19.1
1.19.2
1.19.3
1.19.4
1.19.5
1.19.6
1.19.7
1.19.8