Troubleshooting
Description
Software ver. CVPCA30E_1
Result
Possible Cause
Infusion stops
Air is present in administration
and an alarm is
Set.
activated
The roller or clip on the
administration set is closed
upstream of the pump.
The line was not loaded correctly.
Infusion stops
The roller or clip on the
and an alarm is
administration set is closed
activated
downstream of the pump.
Administration set is kinked.
Cannula is blocked.
Administration set loaded
incorrectly.
NOTE: Operator may change
pressure setting.
An alarm is
Two minutes has elapsed
activated
without a button press during
programming.
Infusion stops
The door of the pump was not
and an alarm is
closed prior to operation
activated
The door has accidentally
opened during operation.
Pump will only
30 Minutes of battery life are
run for another
remaining.
30 minutes if it is
not connected to
mains.
Pump operation
Battery is depleted.
stops. The pump
cannot be used
before being
connected to
mains.
Required Action
Disconnect line from patient and
press STOP/NO key to move to
Pump stopped screen. A long press
(3sec) on the STOP/NO key
followed by a normal press on
STOP/NO key will return to the
priming screen as described on page
25 (through level one code screen).
Open clamp/clip.
Re-load the administration line.
Straighten the set.
Change the Cannula.
Re-load the administration line.
Refer to "Setting Down Occlusion"
section.
Press START/OK to resume.
Close the door of the pump.
Send the pump for service.
Place the pump in charger and
connect to mains.
Place the pump in charger and
connect to mains. Wait 2 minutes
before operating.