Titrating Basal Rate During The Infusion; Patient Activated Bolus - CME BodyGuard 575 Color Vision Operator's Manual

Pca infusion system, pain manager
Table of Contents

Advertisement

12.
Press_
13.
Scroll to select Protocol required then press
14.
Review screen is displayed showing infusion parameters for the protocol
you selected. Check infusion rate, bolus volume, lockout time and 4 hour
limit are correct before proceeding.
NOTE: When MediGuard® is enabled and the max. rate allowable for the
current patients weight has been exceeded than the pump wil alert
"MediGuard® Limit".
NOTE: When selecting a standard pre-programmed protocol ALWAYS
check the review screen carefully to ensure the parameters match the
prescription.
15.
Press
Titrating Basal Rate During Infusion
To change Rate during the infusion, enter the new rate using the keypad and confirm by
pressing the
Rate can be increased up to the "Max Titration Rate" programmed into the specific
infusion. The pump will beep to confirm acceptance of the change and the new rate will
be displayed.
 NOTE: If the desired rate is above the protocol maximum range the pump will not
accept the entry and will display a message notifying the maximum rate
allowed.

Patient Activated Bolus

The BodyGuard® infusion pump can be programmed to allow the patient to request
boluses of a specified volume within predetermined time intervals. The patient can
request the bolus by using the following method:
Software ver. CVPCA30E_1
to confirm.
to commence infusion.
key. Confirm again the new rate by entering the Relevant Code. The
35
.