Priming The Set - CME BodyGuard 575 Color Vision Operator's Manual

Pca infusion system, pain manager
Table of Contents

Advertisement

Priming the Set

1.
Once the set is loaded turn on the BodyGuard®
pump, by pressing
2.
The screen than will prompt for level one code. Enter code 700 and press
3.
From the menu select "Prime" by pressing
Make sure patient is disconnected from the I.V. line
4.
BodyGuard® prompts you to ensure the pump is disconnected from the patient.
WARNING: Ensure the administration set is NOT connected to the patient
during the priming function.
5.
Press
again to commence priming.
6.
A chart will be displayed showing progress versus
total pre-set priming volume. You may stop priming at
any time by pressing
steps if the line is not completely primed after first
priming function.
7.
Make sure all air has been removed from the line and then connect to the patient.
Software ver. CVPCA30E_1
.
or repeat the last two
25
.