Voltage present on internal components may cause severe shock or death upon
contact. Disconnect the Charger from the mains, prior to opening the casing.
Only trained service personnel should open the pump cover.
Blown fuses could cause a fire hazard. Replace blown fuses on the Charger only
with fuses of the same type and rating (see fuse values on the Charger PCB).
Boluses will be infused up to the full volume that is requested unless the user
stops the bolus manually by pressing
. This is true even when the volume
infused is less than the specified Hour Limit for Volume (e.g. 4 Hr Limit) at the
beginning of a bolus and goes over the limit during the bolus.
Make sure the pump is attached securely to the Charger, which is connected
snugly to an IV pole.
A kinked or occluded administration line may impair the operation of the pump
and the accuracy of the infusion. Before operation, verify that the administration
line is not kinked or occluded.
The BodyGuard® should be operated only with the BodySet®/MicroSet®
administration
lines.
Use
of
administration
lines
other
than
the
BodySet®/MicroSet® may impair the operation of the pump and the accuracy of
an infusion.
Drugs must not be administered to the Epidural space unless the drugs are
indicated for this purpose and are administered in accordance with the
indications included in the manufacture's package. Epidural administration of
drugs other than those indicated for Epidural use could result in serious injury to
the patient. For Epidural administration of drugs use MicroSet® only.
Any adjustments, maintenance, or repair of the uncovered pump may impair the
operation of the BodyGuard® Infusion System and/or the accuracy of the
infusion. Any adjustments, maintenance, or repair of the uncovered pump or
charger should be performed by authorized skilled technicians. Any adjustments,
maintenance, or repair of the uncovered pump or charger while connected to the
power should be avoided.
The BodyGuard® Infusion System should be operated within a temperature
range of 18C (59F) to 45C (113F) and up to 85% humidity. Operating the
pump at temperatures and/or humidity other than within that range may affect
accuracy.
Unsafe operation may result from using improper accessories. Use only
accessories and options designed for this system.
Disposables must be compatible with the medicine delivered.
Battery charging is enabled as long as the charger cord is connected to the
mains and the pump is in the charger. Switching the pump off does not
disconnect it from the mains. To disconnect from mains, remove the charger
cord from mains. To disconnect pump from mains, remove it from the charger.
Dropping the BodyGuard® Infusion System could cause damage to components.
If the pump is dropped, return the pump for inspection by qualified service
personnel.