Door Open – Alarm Test
Insert a BodySet® in the pump and close the door. Press the door until a click is
1.
heard.
NOTE: For instructions on how to load and prime an administration set refer to
sections "Loading the Administration Set" & "Priming the Line".
2.
Press
to confirm. Please contact your technician in order to enable the air sensor if
disabled with no need.
3.
Start infusion as instructed in the "Select protocol" section.
4.
Open the pump door.
5.
Door Open alarm will be activated at once.
6.
Close the door and press

Battery Test

1.
Insert a BodySet® in the pump and close the door. Press the door until a click is
heard.
NOTE: For instructions on how to load and prime an administration set refer to
sections "Loading the Administration Set" & "Priming the Line".
2.
Take the pump out of the charger.
3.
Press
to confirm. Please contact your technician in order to enable the air sensor if
disabled with no need.
4.
Start infusion as instructed in the "Select protocol" section.
Software ver. CVPCA30E_1
until the Self-Test screen appears. If air sensor is off, press
.
until the Self-Test screen appears. If air sensor is off, press