Wait untill pump displays "low battery" message
5.
6.
Press
7.
Connect the pump to a charger.
8.
Verify the red icon of battery is lit (on the charger).
9.
Verify the red battery turns green after a minimum of 2 hours.
WARNING: Do not let the pump operate when battery is fully discharged. Pump

Charger Indicator Test

1.
Click the pump into the charger and connect the Charger to mains. Disconnect
the power cord from the AC power outlet.
Verify that the CHARGE indicator is OFF.
2.
3.
Connect the power cord to the AC power outlet. Verify that the CHARGE
indicator is ON.
Software ver. CVPCA30E_1
to turn the pump off.
may turn off during operation on fully discharged battery without
alarm.