Terms Used In Manual

Warning: Indicates that the information is a warning. Warnings advise you of
circumstances that could result in injury or death to the patient or operator.
Read and understand this manual and all warnings completely before
operating the BodyGuard® Infusion System.
Caution: Indicates that the information is a caution. Cautions advise you of
circumstances that could result in damage to the device.
Read and understand this manual and all cautions completely before operating
the BodyGuard® Infusion System.
NOTE: Indicates that the information that follows is additional important
information or a tip that will help you operating the BodyGuard® Infusion
System.

Intended Use

The BodyGuard® infusion Pump system is designed for infusion of medications or fluids
requiring continuous or intermittent delivery at precisely-controlled infusion rates
through
clinically
subcutaneous, percutaneous, intra-arterial, PCA, enteral, in close proximity to nerves,
and into an intraoperative site (soft tissue/body cavity/surgical wound site). The system
is intended for patients who require maintenance medications, analgesics, PCA therapy,
parenteral and enteral nutrition fluids, chemotherapeutic agents and general fluids
therapy in hospital and home care environments.
The BodyGuard® 575 PCA Infusion Pump is designed for patients who require PCA
pain management. We recommend dedicated pumps for differentiation PCA
administration to prevent medication errors. The BodyGuard® is available in a color-
coded PCA format. Please ask your local representative or contact us direct for further
details. Please ensure the pumps are only used by, or under the supervision of, trained
medical staff.

Warnings

To avoid possible personal injury or loss of life, observe the following:
Using the pump with air in line detector off may cause an embolism resulting
death or paralysis. Switching the air detector off is not recommended and should
be used only when absolutely necessary. Please contact your technician in order
to enable the air sensor if disabled with no need.
Read the entire Operation Manual before using the pump, since the text includes
important precautions.
The maximum volume that may be infused under SINGLE FAULT CONDITION
is 0.1 ml
Software ver. CVPCA30E_1
acceptable
routes
of
administration,
11
including
intravenous,