BodyGuard® Default Settings
The pump comes programmed with a number of default parameters that can be
viewed using 'Review Set Up' from the main menu and changed using the
technician Access Code in 'Change Set Up' mode.
PARAMETER
Battery Level
Lock Out
Time, Date
Tube Temperature
Alarm Volume
Pressure Level
Priming Volume
Priming Rate
Bolus Rate
Maximal Basal Rate
KVO Rate
Program Limits
Air in line limit
Indicator Led
Set Time & Date
Event Log
Serial No.
Production Date
Operation Time
Software Ver.
Volume Calibration
Pressure Delta
Pressure CAP
Exit/Restart
 NOTE: Only highlighted parameters can be changed (see 'Change Set Up')
Software ver. CVPCA30E_1
PURPOSE
Current battery charge
Shows whether access code
security is on or off.
Shows the temperature of the
tubing set
Volume of alarms/alerts
Occlusion Pressure Setting
Amount of fluid delivered during
set priming.
Rate at which priming occurs.
Speed at which boluses are
delivered.
Maximum continuous basal rate.
Keep Vein open rate.
Limits in ml, mg, µg or in bolus
Sets the maximum air bubble size
needed to cause an alarm
Switches the LED indicator
On & Off
History Records the last 500 pump
programming events
Serial Number Unique Pump
Date of manufacture
Hours since last service
Pump software version
Volumetric calibration value
Pressure calibration Delta value
Pressure calibration CAP value
Exit/Restart when OK
47
RANGE
Empty >> Full
ON/OFF
C
3-7
250mmHg – 520mmHg
(5psi – 10psi )
0-20ml
Default 4 ml
0-1200 ml/hr.
Default 600 ml/hr
According to back
pressure in system – max.
rate 30ml/h
0.1-0.5 ml/hr
Off, 1, 4, or 24 hours.
Off when set to 0.
ON/OFF
Fixed data
Fixed data
Week, month, year
Fixed data
Fixed data
Fixed data – 255